Product model |
Part name and model |
AR-960-I |
|
|
X-ray generation device |
High-pressure trigger |
FSYG60 |
1 |
|
X-ray tube |
LQ16-XA4 |
1 |
|
|
High-voltage cable |
GD-1 |
2 |
|
|
X-ray imaging system |
Flat-panel detector |
DR 17/17 |
1 |
|
DR 14/17 |
/ |
||
|
Computer |
DR-1000 |
1 |
|
|
Display unit |
XS-1000 |
1 |
|
|
Software systems |
DR-1000 |
1 |
|
|
Subsidiary equipment |
Photo bed |
SYC-2 |
1 |
|
Chest support |
SPJ-2 |
1 |
|
|
Beaming device |
XSQ-4 |
1 |
|
|
Professional medical image processing software |
|||
1. Quick imaging:Imaging time below 6S to meet large-scale medical examination requirements.
2. Image replay:Image thumbnail display, sequence playback tool, digital image replay.
3. Image processing:Window width/window position adjustment, arrow, text and other labels, angle, distance measurement, image zoom, pan, left and right flip, up and down, rotation, black and white inversion, shadow screen selection.
4. Image storage:Image real-time storage, DICOM image sending, disc burning, export storage (optional in a variety of storage formats: Bitmap, Jepg, AVI and other office software directly used in Word and PowerPoint, convenient for doctors to diagnose reports and paper writing.
5. DICOM 3.0:Perfect compatibility with brand laser cameras and PACS systems.
6. Medical records management:Database management, graphic reporting, support for WORKLIST
Product performance and parameters:
|
Project |
Product performance and parameters |
||
|
X-ray generation device |
High frequency reversal frequency: no less than 120kHz |
||
|
Maximum output power: Photo: a) P should be 100KV, 500mA, 50kW (large focus) b) P should be 110KV, 160mA, 17.6KW (small focus) |
|||
|
X-ray tube voltage: tube voltage regulation range: 40 kV to 150 kV, 1 kV per step |
|||
|
X-ray tube current: photographic tube current regulation range: 10mA-630mA, split adjustment mode, selected by national standard: GB9706 r10 coefficient |
|||
|
Tube current: tube current product adjustment range: 1mAs-1000mA, the current product using split a djustment mode, according to national standards: GB9706 r20 coefficient selection; |
|||
|
Anti-overload: X-ray machine has anti-overload measures |
|||
|
Photo preparation time: Photo preparation time should not be greater than 1s |
|||
|
X-ray imaging system |
Network communications: DICOM 3.0 compliant |
||
|
Amorphous silicon flat panel detector |
DR 17/17 |
No less than 430 mm x 430 mm (17 in x 17 inches) |
|
|
Amorphous silicon full board seamless, ultra-thin design, thickness and box size of 15mm |
|||
|
Image spatial resolution should be no less than 3.6 LP/mm |
|||
|
Software: user management, capture images, exposure, image preview, image processing, image measurement, image printing |
|||
|
Subsidiary equipment |
Photo bed |
Photo bed surface, left and right range is not less than 900mm, front and rear range of motion is not less than 320mm |
|
|
Ball tube can be used left and right movement stroke should not be less than 1200mm, up and down movement trip should not be less than 1000mm |
|||
|
The Focus distance of the X-ray tube to the flat plate detector is no less than 1200mm |
|||
|
The X-ray tube assembly should rotate no less than 100 degrees around its axis and the arm shaft should not rotate less than .90 degrees |
|||