Technical Specifications
|
PARAMETER |
SPECIFICATION |
|
Trolley type full digital color Doppler ultrasonic mainframe |
|
|
Ultrasonic host operating system |
Windows operating system |
|
Applications |
Abdomen, obstetrics, gynecology, heart, urinary system, small organs, superficial, blood vessels, pediatrics, newborns, musculoskeletal |
|
Probes |
Convex probe, Tran-vaginal probe, Linear probe, Micro-convex probe, Cardiac probe,4D Volume probe |
|
Applications and report |
Support abdomen, obstetrics, gynecology, heart, urology, small organs, superficial, blood vessels,pediatrics, etc. advanced measurement software package, report software package, patient management software package |
|
carotid artery intima measurement thickness(IMT) |
|
|
Automatic spectral envelope measurement |
|
|
Full digital transmission and reception of beam synthesizer |
|
|
Color Doppler imaging(C) |
|
|
Pulse Doppler Imaging(PW) |
|
|
Coherent Contrast imaging(CCI) |
|
|
Continuous wave Doppler imaging(CW) |
|
|
B/C/D Real-time three synchronous imaging |
|
|
Power Doppler imaging(PDI) |
|
|
Direct power Doppler imaging(DPDI) |
|
|
M mode imaging |
|
|
Anatomic M mode imaging |
|
|
Color Doppler M mode imaging |
|
|
Elastography |
|
|
Tissue Doppler imaging(TDI) |
|
|
Strain rate imaging (SRI) |
|
|
Tissue harmonic imaging(THI) |
|
|
Fusion harmonic imaging(FHI) |
|
|
Speckle Reduce imaging(SRI) |
|
|
Panoramic imaging |
|
|
Deflection imaging |
|
|
Trapezoidal imaging |
|
|
Adaptive velocity optimization |
|
|
Free hand 3D |
|
|
Real time 3D imaging(3D/4D) |
|
|
DICOM3.0 |
|
|
Monitor |
≥21.5 inch,high definition ultrasonic display |
|
≥10.4 inch touch screen |
|
|
Physical clipboard |
save the image on the left side of the screen, which can be directly saved or deleted. |
|
The system has the function of on-the-spot upgrade |
|
|
Presupposition |
for different inspection of the viscera, preset the inspection conditions for the best image, reduce the adjustment of the operation, and the commonly used external adjustment and combination regulation. |
|
Probe interface |
4 |
|
Chinese and English System, Chinese and English input, optional |
|
|
Depth |
≥360mm |
|
Extended imaging |
|
|
Probes |
|
|
Convex probe |
Fundamental Frequency: 2.0MHz/2.3MHz/2.5MHz/3.0MHz/3.5MHz/4.0MHz/4.6MHz/5.0MHz/5.4MHz, Harmonic Frequency: 4.0MHz/4.6MHz/5.0MHz |
|
Linear probe |
Fundamental Frequency: 4.0MHz/4.6MHz/5.0MHz/6.0MHz/7.0MHz/8.0MHz/9.2MHz/10.0MHz/12.0MHz/13.3MHz, Harmonic Frequency: 8.0MHz/9.2MHz/10.0MHz |
|
Trans-vaginal probe |
Fundamental Frequency: 3.0MHz/3.5MHz/4.0MHz/5.0MHz/5.4MHz/6.0MHz/7.0MHz/8.0MHz/10.0MHz Harmonic Frequency: 6.0MHz/7.0MHz/8.0MHz |
|
Micro-convex probe |
Fundamental Frequency: 3.0MHz/3.5MHz/4.0MHz/5.0MHz/5.4MHz/6.0MHz/7.0MHz/8.0MHz, Harmonic Frequency: 6.0MHz/7.0MHz/8.0MHz |
|
Cardiac probe |
Fundamental Frequency: 1.7MHz/1.9MHz/2.1MHz/2.5MHz/3.0MHz/3.4MHz/3.8MHz/4.2MHz/5.0MHz, Harmonic Frequency: 3.4MHz/3.8MHz/4.2MHz |
|
4D Volume probe |
Fundamental Frequency: 2.0MHz/2.5MHz/3.0MHz/3.3MHz/3.7MHz/4.0MHz/5.0MHz/6.0MHz, Harmonic Frequency: 4.0MHz/5.0MHz/6.0MHz |
|
2D imaging mode |
|
|
Gain |
0-100,Step 2 adjustable |
|
TGC |
8 segment adjustable |
|
Maximum focus point |
≥7, which can be moved throughout the whole process |
|
Speckle reduction |
0-5,5 level |
|
Space Synthesis |
0-2,2 level(Liner probe: 3 level, cardiac probe:0) |
|
Dynamic |
30-180,35 level,step 5 adjustable |
|
Line density |
low、middle、high,3 level |
|
Frame correlation |
0-4,4 level |
|
Noise reduction |
0-5,5 level |
|
Edge Enhancement |
0-5,5 level |
|
Sound power |
2-10, 9 level |
|
Grey scale |
0-67, 67 level |
|
False color |
0-67,67 level |
|
Image style |
Soft-Comparison,2 level |
|
The screen has real-time display of voice power, probe frequency, dynamic range, pseudo color, gray scale and other 11 parameters can be adjusted |
|
|
Color Doppler imaging mode |
|
|
Blood gain |
0-100,Step 2 |
|
Parameter display |
Velocity、Variance |
|
B-Restrain(B/W restrain) |
0-7, 7 level |
|
Speed Through |
0-8, 8 level |
|
Sampling number |
6-24, 7 level |
|
Blood flow preferred |
0-8, 8 level |
|
Filtering |
1-6, 6 level |
|
Sound power |
2-6, 4 level |
|
Noise reduction |
0-4, 4 level |
|
Smooth treatment |
0-4, 4 level |
|
Frame correlation |
0-6, 6 level |
|
Chromatography(Blood flow graph) |
0-37, 37 level |
|
Line density |
Low-Middle-High, 3 level |
|
Frequency |
4 level adjustable |
|
Velocity:Minimum 0.4K,Maximum 40.5K Convex probe:0.4K-4.3K-38.5K Linear probe:0.4K-14.7K-39.0K Trans-vaginal probe:0.4K-7.8K-39.7K Volume probe:0.4K-4.2K-34.8K Micro-convex probe:0.4K-10.3K-40.5K cardiac probe:0.4K-7.8K-39.7K |
|
|
PS: The frequency of the probe changes and the frequency value changes |
|
|
PS: Frame rate changes with speed |
|
|
Pulse wave Doppler (PW) |
|
|
Gain |
0-100,Step 2 |
|
Spectrum envelope function |
real time automatic spectrum envelope, manual spectrum envelope, and other modes. The system automatically analyses and displays various data such as PSV, EDV, RI, PI, S/D, ACC, HR and so on. Can wake up or close |
|
Sample volume |
0.5mm~30mm |
|
Blood angel |
-75—75 degree,Step 5 |
|
False color |
0-67, 67 level |
|
Dynamic range |
20-40, 4 level |
|
Filter |
0-9, 9 level |
|
Smooth treatment |
1-4, 4 level |
|
Sound power |
2-5, 4 level |
|
Volume |
0-100, 10 level,Step 10 |
|
Audio filtering |
0-4, 4 level |
|
Base line |
-1.0~1.0 |
|
Grey map |
0-67, 67 level |
|
Scan velocity |
100-500, 6 level |
|
PRF:Minimum 0.5K,Maximum 87.5K Convex probe:0.5K-4.3K-63.3K Linear probe:0.5K-14.5K-78.4K Trans-vaginal probe:0.5K-8.1K-78.4K Volume probe:0.5K-4.2K-53.8K Micro-convex probe:0.5K-10.3K-81.1K cardiac probe:0.5K-4.3K-87.5K |
|
|
Frequency |
4 level |
|
PS: The frequency of the probe changes and the PRF value changes |
|
|
PS: The frequency of the probe changes and the frequency value changes |
|
|
Continuous Wave Doppler (CW) |
|
|
Support probe |
Cardiac probe |
|
Adjustment of B mode parameters is switchable |
|
|
Gain |
0-100,Step 2 |
|
Sampling line position is adjustable |
|
|
PRF |
0.9K~36.1K |
|
Baseline |
-1.0~1.0 |
|
Blood angel |
-75~75 degree |
|
Grey map |
0-67 |
|
Scan velocity |
100-300 |
|
False color |
0-67 |
|
Dynamic range |
20-40 |
|
Filtering |
0-9,9 level |
|
Smooth treatment |
1-4 |
|
Frequency |
2.0MHz/2.3MHz/2.5MHz/3.0MHz,4 level adjustable |
|
Sound power |
2-5 |
|
Volume |
0-100 |
|
Audio Filtering |
0-4 |
|
Anatomical M imaging |
|
|
Support probe |
Convex probe, Linear probe, Cardiac probe |
|
Adjustment of B mode parameters is switchable |
|
|
Gain |
0-100,Step 2 |
|
M Sampling line angel is adjustable |
|
|
M Sampling line length is adjustable |
|
|
Sampling line |
3,Can be displayed or hidden separately |
|
Blood flow M model (MC) |
|
|
Adjustment of B mode parameters is switchable |
|
|
Gain |
0-100,Step2 |
|
MC Sampling line angel is adjustable |
|
|
MC Sampling line length is adjustable |
|
|
Frequency |
4 level |
|
Sampling number |
6-24 |
|
Speed through |
0-8, 8 level |
|
Scan velocity |
150-500 |
|
Frame correlation |
0-6, 6 level |
|
Filtering |
1-6,6 level |
|
Blood flow preferred |
0-8,8 level |
|
Smooth treatment |
0-4,4 level |
|
Map |
0-37, 37 level |
|
Elastography |
|
|
Adjustment of B mode parameters is switchable |
|
|
Gain |
0-100,Step 2 |
|
B/E, Double real-time display on the same screen |
|
|
Probe displacement curve display |
Up/Down |
|
Pressure indicator bar display |
|
|
Frequency |
8-9 level,Adjustable;According to the probe display |
|
Noise reduction |
0-2, 2 level |
|
Frame correlation |
0-3, 3 level |
|
Comparison |
0-13, 13 level |
|
False color |
0-3, 3 level |
|
Don’t support cardiac probe |
|
|
Tissue Doppler imaging (TDI) |
|
|
Support probe |
Cardiac probe |
|
Adjustment of B mode parameters is switchable |
|
|
Gain |
0-100,step 2 |
|
ROI area adjustable |
|
|
Sampling number |
6-24 |
|
Velocity |
0.4K-8.0K |
|
Frame correlation |
0-6,6 level |
|
Tissue preferred |
0-7, 7 level |
|
Frequency |
2.0MHz/2.3MHz/2.5MHz/3.0MHz |
|
Support color reversal |
|
|
Strain rate imaging |
|
|
Support probe |
Cardiac probe |
|
Adjustment of B mode parameters is switchable |
|
|
ROI area adjustable |
|
|
Gain |
0-100,Step 2 |
|
Sampling number |
6-24,6 level |
|
Axial average |
1-4, 4 level |
|
Velocity |
0.4K-8K |
|
Frame correlation |
0-6, 6 level |
|
Tissue optimization |
0-7,7 level |
|
Panoramic imaging |
|
|
Support probe |
Linear probe |
|
Speckle Reduction |
0-5, 5 level |
|
Deflection imaging |
|
|
Support probe |
Linear probe |
|
Adjustment of B mode parameters is switchable |
|
|
Deflection angel |
8 level |
|
Speckle reduction |
0-5, 5 level |
|
Dynamic rate |
30-180,Step 5 |
|
Line density |
low-middle-high,3 level |
|
Frame Correlation |
0-4, 4 level |
|
False color |
0-67, 67 level |
|
Image style |
Soft-Comparison,2 level |
|
Noise reduction |
0-5, 5 level |
|
Edge Enhancement |
0-5,5 level |
|
Sound power |
2-10,8 level |
|
Grey map |
0-67,67 level |
|
Trapezoidal imaging |
|
|
Probe support |
linear probe |
|
Adjustment of B mode parameters is switchable |
|
|
Deflection angel |
8 level |
|
Speckle reduction |
0-5, 5 level |
|
Dynamic rate |
30-180,Step 5 |
|
Line density |
low-middle-high,3 level |
|
Frame Correlation |
0-4,4 level |
|
False color |
0-67, 67 level |
|
Image style |
Soft-Comparison,2 level |
|
Noise reduction |
0-5, 5 level |
|
Edge Enhancement |
0-5,5 level |
|
Sound power |
2-10,8 level |
|
Grey map |
0-67, 67 level |
|
Space Synthesis |
0-2, 2 level |
|
Freehand 3D imaging |
|
|
Support probe |
convex probe, linear probe |
|
Display model |
4 pictures |
|
Image Rotation X/Y/Z Axis |
|
|
Multi-slice Visibility |
|
|
Real-time 4D imaging |
|
|
Support probe |
4D volume probe |
|
Adjustment of B mode parameters is switchable |
|
|
Gain |
0-100,Step 2 |
|
Display model |
one image、two images、four images |
|
Image Rotation |
X/Y/Z Axis |
|
Multi-slice Visibility |
|
|
Light&Shade inversion |
|
|
Smooth |
0-4, 4 level |
|
Threshold level |
0-129, Step 3 |
|
Transparency |
1-509,Step 10 |
|
Render type |
4 kinds,Surface、maximum、minimum、perspective |
|
Extended Imaging |
|
|
Gain |
0-100,Step 2 |
|
TGC |
8 segment adjustable |
|
Maximum focus point |
≥7, which can be moved throughout the whole process |
|
Speckle reduction |
0-5,5 level |
|
Space Compound |
0-2,2 level(Linear probe: 3 level,don’t support cardiac probe) |
|
Dynamic range |
30-180,35 level,Step 5 |
|
Line density |
Low、Middle、High,3 level |
|
Frame correlation |
0-4,4 level |
|
Noise reduction |
0-5,5 level |
|
Edge enhancement |
0-5,5 level |
|
Sound power |
2-10, 9 level |
|
Grey map |
0-67, 67 level |
|
False color |
0-67,67 level |
|
Image style |
Soft-Comparison,2 level |
|
Extended level: Maximum 72 level Convex probe:9 level Trans-vaginal probe:72 level Micro-convex probe:29 level Cardiac probe:40 level 4D Volume probe:17 level |
|
|
PS:The screen has real-time display of voice power, probe frequency, dynamic range, pseudo color, gray scale and other 11 parameters can be adjusted |
|
|
PS: When the probe scan range reaches the maximum, the space synthesized is 0. |
|
|
Measurement and analysis function |
|
|
General measurement: Distance, area, ellipse, cross line, angle, distance ratio, volume, Volume (ellipse), area ratio, diameter, joint angle |
|
|
Cardiac |
Automatic spectrum envelope、LV、Main Pulmonary artery diameter、RVEDd、RVEDs、LVM、LAV、HR、MVF、AO、AR、LVOT、TVF、Pulmonic valve、Pulmonary vein、RV、Doppler fetal heart sound、LVET、LVM、LVMI、AV |
|
Vascular |
carotid intima (IMT), length stenosis ratio, area stenosis ratio, IMT (back wall), IMT (front wall) |
|
OB |
Fetal routine、AFI、TW、GS、CRl、OFD、HL、ulna、NT、Fibula、Nbonel、Radial、Tibia |
|
GYN |
uterus、cervix、corpus uteri/cervix uterus、left ovarian vein、right ovarian vein、dominant follicle、intima thickness |
|
Urology |
prostate、residual urine、left kidney、right kidney、left suprarenal vein、right suprarenal、left testis、right testis、left seminal vesicle、right seminal vesicle |
|
Abdomen |
liver、CHD、partal vein diameter、cholecyst、CBD、pancreas、spleen、Internal diameter of abdominal aorta、kidney |
|
Small parts |
Thyroid |
|
Software package |
Measurement package、Software package、Medical records management software package |
|
Graphic and text management system |
|
|
Host build in 2 hard disk(SSD 120+1T),Start fast and stable |
|
|
Movie playback |
≥1200 frames |
|
Internal file information management system: can record patient number, name, check number, check date and so on, and can be searched and managed by numbering, checking number, name and so on. |
|
|
Type of report |
16 |
|
One key fast report graphic and text management |
|
|
Interface |
|
|
USB interface |
4 |
|
HDMI interface |
1 |
|
RJ-45 interface |
1 |
|
Grounding wire interface |
1 |
|
DVD RW |
1 |
|
Configuration |
|
|
Trolley type full digital color Doppler ultrasound diagnostic system |
|
|
Probe |
convex array probe (standard), linear probe (optional), Trans-vaginal probe (optional), cardiac probe (optional), 4D volume probe (optional) |
|
≥13 quick adjusting knobs |
|
|
Technology, after-sales service and other requirements |
|
|
After acceptance, the warranty is free for two years(Provide manufacturer warranty certificate) |
|
|
Manufacturer has ISO13485 certification and EU CE certification. |
|