S1100 ICU Ventilator

Application

The ventilator makes a good performance in operation room, ICU department and emergency treatment. It used to assist or replace the spontaneous breathing for adult pediatric and neonatal more than 2kg.Due to the flexible configuration, good quality and competitive price, S1100 has soon become the superstar of market.

Features

1.15” TFT LCD screen displays the Ventilation parameters, Alarming information and oscillograms, make every operation more easily.

Multiple ventilation modes, suitable for ICU, emergency department and operation room etc:(PRVC,APRV,DUOLEVEL,V-SIMV,P-SIMV,IPPV,A/C,PCV,PSV,SPONT/CPAP, SIGH, MANUAL)

2.5 oscillograms for your choice, 3 of them can be displayed on the screen at the same time.

3.Humidifier can heat and wet breathing gas,makes it comfortable for patient to breathe

4.Rapid oxygen supply, automatically offer high flow rate oxygen within two minutes.

5.Nebulization make medicine into small liquid and particle, easier and quicker for patient to breathe in.

6.High temperature resistance breathing circuit is reusable and anti-pollution.

Safety

1.14 types of sound and visual alarm information, easier for users to do some error checking and troubleshooting.

2.Built-in oxygen concentration sensor, ensure stable precision of oxygen concentration

3.Easy to move with four casters, easy to stop with two brakes

4.Separate design of electronic circuit and gas flow rate keep safe running of ventilator.

5.Compact long life internal battery can provide emergency power, avoid risk of patient

6.Self-check before operation, eliminate system mistake

Technical Specifications

NAME

PARAMETER

Ventilation modes

PRVC,APRV,DUOLEVEL,V-SIMV,P-SIMV,IPPV,A/C,PCV,PSV,SPONT/CPAP, SIGH, MANUAL

Ventilator parameter ranges

Tidal volume(Vt)

0~2000mL

Frequency (Freq)

1min~100min

Oxygen concentration

21%-100%

I/E

4:1~1:8

PEEP

0cmH2O~ 40 cmH2O

Pressure Triggering Sensitivity (Ptr)

-20cmH2O~20cmH2O (Based on PEEP)

Flow Triggering Sensitivity (Ftr)

0.5L/min~30L/min

Pressure Control

5cmH2O~80cmH2O

Pressure Support

0cmH2O~80cmH2O

SIGH

0(off) 1/100~5/100

Apnea Ventilation

OFF, 5s~60s

Pressure Limit

20cmH2O~100cmH2O

Monitored Parameters

Frequency (Freq)

0 /min~ 100 /min

Tidal volume(Vt)

0 mL~ 2500 mL

MV

0L/min~99L/min

Airway pressure

0cmH2O~100cmH2O

Dynamic Lung Compliance Testing

1mL/cmH2O~1000mL/cmH2O

Oxygen Concentration

15%-100%

Alarm and protection  

The AC power failure alarm

Power failure or no connection

Internal battery backup low voltage alarm

11.3±0.3V

No tidal volume

No tidal volume within 6s

High Minute Volume alarm

Low Minute Volume alarm

5L/min-99L/min

1L/min-30L/min

High Airway pressure alarm

Low Airway pressure alarm

High oxygen concentration alarm

Low oxygen concentration alarm

20cmH2O-100cmH2O

0cmH2O-20cmH2O

19%-100%

18%-99%

Continuous pressure alarm

(PEEP+1.5cmH2O) over 16s

Suffocation warning

Fan error

Oxygen deficit

5-60s

Show on screen

Show on screen

The maximum limited pressure

<12.5 kPa

Working conditions

Gas source

O2,AIR

Pressure

280kPa-600kPa

Voltage

220V±22V

Power frequency

50Hz±1Hz

Input power

900VA(with air compressor)

250VA(without air compressor)

Oscillogram

P-T

Pressure Time

F-T

Flow-Time

V-T

Volume Time

P-V Loop

Pressure-Volume Loop

F-V Loop

Flow Volume Loop

P-F Loop

Pressure-Flow Loop