AR-201 Ventilator

Technical Specifications

PARAMETER

SPECIFICATION

Display Mode

High-definition 5.7” LCD screen display

Ventilation Mode

IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH

Ventilation Parameters

Tidal Volume

50~1500ml

Rate

2~99bpm

SIMV Rate

2~20bpm

I:E

2:1~1:8

Inspiratory Trigger Pressure

-1.0~2.0kPa

PEEP

0~2.0kPa

Pressure Range

1.0~6.0kPa

SIGH

1.5 times the inspiratory time

Parameters for Ventilation Monitoring

Tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate – time waveform, PEEP, inspiratory trigger pressure.

Power

AC 220V±10% 50Hz±1Hz; UPS, Storage battery

Mechanical Arms

SECURITY ALARM SYSTEM

Power Alarm

Airway Pressure Alarm

Upper Limit Setting Range

1.0~6.0kPa

Low Limit Setting Range

0.4~2.0kPa

Per-minute Ventilation Volume Alarm

Upper Limit Setting Range

3.0~30L/min

Low Limit Setting Range

1.0~10L/min

Sustained High-pressure Alarm

It will give alarm when stress have consistently been higher than 2.5kPa.

Suffocation Alarm

It will give sound and light alarm if there is no tidal volume input for 15 seconds.