High-strength engineering plastic rack, light, beautiful and corrosion resistant
Adult, child
O2: 0.27--0.55Mpa;
N2O: 0.27--0.55MPa.
O2: 0.1--1.0L/min; 1.1--10L/min
N20: 0.1--1.0L/min; 1.1--10L/min
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure < 0.2 MPa, the flow of nitrous oxide would be cut off.
25--75L/min
Low Oxygen Pressure Alarm
There will be sound alarm when the oxygen pressure < 0.2MPa
It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0-5vol%.
Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.
Working mode: all-close, semi-close, semi-open;
Releasing pressure: 0.5-7kPa.
Bellows for adults, bellows for children, tidal volume range:0-1500 ml.
10.4” Touch Screen, High-definition TFT color LCD.
IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH (IMV, VCV, PCV)
Rate 1--99bpm
SIMV rate 2--20bpm
I:E 2:1--1:8 (4:1--1:8)
Inspiratory trigger pressure -1.0--2.0kPa
PEEP 0--2.0kPa
Pressure range 1.0--6.0kPa
SIGH 1.5 times the inspiratory time (70-150 adjustable)
Inspiratory platform 0--50%
Tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate – time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform
(autonomous inspiratory rate, lung compliance, airway resistance, flow-volume loop, pressure-volume loop)
21%--100%
Upper limit setting range 21%--100%;
Low limit setting range 10%--80%.
Low limit setting range 0.2--5.0kPa.
Per-minute Ventilation Volume Alarm
Upper limit setting range 3.0--30L/min;
Low limit setting range 1.0--10L/min.
Sustained High-pressure Alarm
It will give alarm when stress have consistently been higher than 2.5 kPa
It will give sound and light alarm if there is no tidal volume input for 15 seconds
AC 220V±10%, 50Hz±1Hz, (Storage battery, UPS)
Mechanical Arms (End-expiratory CO2 monitoring)
Note: Items within brackets are optional.