Technical Specifications
Main Body
High-strength engineering plastic rack, light, beautiful and corrosion resistant
Scope of Application
Adult, child
Gas Source
O2: 0.27--0.55Mpa;
N2O: 0.27--0.55MPa.
Flow Meter
O2: 0.1--1.0L/min; 1.1--10L/min
N20: 0.1--1.0L/min; 1.1--10L/min
Air: 0.1--10L/min
O2, N2O Linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure < 0.2KPa, the flow of nitrous oxide would be cut off.
Flow Rate of Rapid Oxygen Supply
25--75L/min
Low Oxygen Pressure Alarm
There will be sound alarm when the oxygen pressure < 0.2MPa
Vaporizer
It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0-5vol%.
Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.
Respiratory Circuit
Working mode: all-close, semi-close, semi-open;
APL: 0.5-7kPa
Respiratory Bellows
Bellows for adults, tidal volume range:0--1500ml;
Bellows for children, tidal volume range:0-300ml.
Anyone can be chosen for application as required by the customer
VENTILATOR
Display Mode
10.4” Touch Screen, High-definition TFT color LCD
Ventilation Mode
IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH, IMV, VCV, PCV
Ventilation Parameters
Tidal volume 20--1500ml
Rate 1--99bpm
SIMV rate 2--20bpm
I:E 4:1--1:8
Inspiratory trigger pressure -1.0--2.0kPa
PEEP 0--2.0kPa
Pressure range 1.0--6.0kPa
SIGH 1.5 times the inspiratory time, 70-150 adjustable
Inspiratory platform 0--50%
Parameters for Ventilation Monitoring
Tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure – time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform, autonomous inspiratory rate, lung compliance, airway resistance, flow-volume loop, pressure-volume loop
Monitoring of Oxygen Concentration
21%--100%
SECURITY ALARM SYSTEM
Oxygen Concentration Alarm
Upper limit setting range 21%--100%;
Low limit setting range 10%--80%
Airway Pressure Alarm
Upper limit setting range 0.3--6.0 kPa;
Low limit setting range 0.2--5.0 kPa.
Per-minute Ventilation Volume Alarm
Upper limit setting range 3.0--30L/min;
Low limit setting range 1.0--10L/min.
Sustained High-pressure Alarm
It will give alarm when stress have consistently been higher than 2.5 kPa
Suffocation Alarm
It will give sound and light alarm if there is no tidal volume input for 15 seconds
POWER ALARM
Power
AC 220V±10%, 50Hz±1Hz, Storage battery (UPS)
Mechanical Arms (End-expiratory CO2 monitoring)
Note: Items within brackets are optional.
M-9000E Patient Monitor
12 inch Six Standard parameters: ECG, RESP, NIBP, SPO2, 2_TEMP, PR/HR