Features
1. Compact in size, easy and direct operation, portable
2. Accurate pace pulse identification and automatic analysis function.
3. Four working modes: Manual / Auto / Physical Examination / Storage
4. 80mm, 3 channel format recording
5. Freeze function: can replay 12 seconds cardiograph front and back
6. Fast ECG test under the [Physical Examination], saving power and paper
7. Store and replay cardiograph without printing parameters and diagnosis under the Storage Mode.
8. 5''800×480 graphic 5 inch color screen
9. Alphabet keyboard fast inputting clinical information
10. Adapt to 110-230V, 50/60Hz AC power supply. Built in rechargeable Li-ion battery.
11. 12 lead waveforms view with 250 patient information storage (Extendable)
12. Built-in USB, RS232 interface, EXGNET software optional
13. USB interface for net transmission, USB flash disk and USB printer.
Technical Specifications
|
PARAMETER |
SPECIFICATION |
|
Lead |
Standard 12 leads |
|
Lead acquisition |
12bit/1000Hz(synchronously 12 leads) |
|
Work mode |
MAN, AUTO, ANALYSIS Modes |
|
Filter |
AC Filter : 50Hz / 60Hz EMG Filter : 25Hz / 45Hz Anti-Drift Filter : 0.15Hz (Adaptive) |
|
CMMR |
≥100dB (with AC FILTER) |
|
Input circuit |
Floating; Protection circuit against Defibrillator effect |
|
Input Impedance |
≥50MΩ |
|
Input circuit current |
≤0.05μA |
|
Patient current leakage |
<10μA |
|
Calibrating voltage |
1mV±2% |
|
Voltage tolerance |
≥±500mV |
|
Time constant |
≥3.2s |
|
Frequency response |
0.05~160Hz(-3db) |
|
Noise level |
<15μVp-p |
|
Inter-channel Interference |
≤0.5mm |
|
Sensitivity |
5,10,20, 40, Auto (mm/mV) |
|
Recording mode |
Default Man. Mode is 3 ch format Default Auto. Mode is 3 ch+ format |
|
Paper speed |
6.25、12.5、25、50mm/s(±3%) |
|
Paper |
80mm *20m Roll paper |
|
Communication |
RS232/ USB port |
|
Power Supply |
AC 110-230V( ±10%),50/60Hz ( ±1Hz),60VA DC Built-in rechargeable battery,14.4V(2000mmAh) |
|
Dimension/Weight |
300mm×230mm×72mm/2.8Kg |