Features
1. No electricity will be needed,only manual operation
2. Module rack with polyester medical material, high strength and light weight.
3. Superior thin spray technology, delicate color, beautiful and durable.
4. High precious vaporizer, Halothane,Enflurane, Isofurane and Sevoflurane could be choosed freely
5. Main unit are only 2Kg(w/o vaporizer), light weight and easy to move
6. With a dedicated handle, could move by hand-pull.
7. Two gas source:O2 ,N2O, suitable for more patient type
8. Classical CO2 circle absorber with 1 canister,low resistance,minimum gas leakage, 60cmH2O peak pressure solution.
9. Ventilator (Optional):TH-1,TH-1A
Technical Specifications
|
PARAMETER |
SPECIFICATION |
|
Ventilation mode |
Closed, Semi-closed, Semi-open |
|
Application |
2-100KG animals |
|
Gas source |
O2, N2O 0.25Mpa~0.65Mpa |
|
Two tube (O2, N2O) |
Scope,0—10L/min.With linkage device,O2 concentration not low than 25% |
|
O2 flush |
more than 25L/min |
|
Respiratory mode |
Manual,Oxygen |
|
TV |
50~2000mL |
|
I:E |
according to the doctor operation |
|
BPM |
according to the doctor operation |
|
APL Valve pressure |
0.5-6KPa |
|
Monitoring |
Airway pressure,respiratory mode |
|
Three Gauges |
one for O2 ,one for N2O, one for airway pressure |
|
Environment Requirement |
|
|
Temperature |
Operation: 18~36 ºC |
|
Storage: -20~55 ºC |
|
|
Humidifier |
Operation:45%~75% |
|
Storage:≤93% |
|
|
Atmospheric pressure |
Operation:0.95~1.05 bar |
|
Storage: 500hPa~1060hPa |
|
|
Altitude |
Operation:500 to 800mmHg (3565~-440m) |
|
Storage:375 to 800mmHg (5860~-440m) |
|
TH-1 Ventilator (Optional)
1. Control method Gas driving electronic controlled
2. Display High light LED display
3. Ventilation mode Volume mode and Manual mode
4. Tidal Volume 50 ~ 1600ml
5. Breath frequency 1 ~ 90BPM
6. I:E ratio 4:1 ~ 1:6
7. Monitor parameter Tidal volume, breath frequency, MV
8. Alarm MV high alarm, MV low alarm, TV high alarm, TV low alarm